HORIBA Micros ES 60

Overview

  • Highly compact and lightweight
  • Micro-sampling on whole blood: 10μL
  • Precise cellular identification through electronic impedance variation method
  • Haemoglobin measurement with photometry

Features

ABX Micros ES 60

  • Integrated barcode reader (identification of patient samples, QC and reagents)
  • Integrated thermal printer (full data printout, including PLT, RBC and WBC curves)
  • User-friendly interface (colour touch screen, virtual keyboard)
  • Data transfer with USB key (archiving to PC, upload of QC target values)
  • Unique reagent pack system

Specifications

Technical

Weight

30.9 lb / 16 kg

Power supply

100 V to 240 V – 50 Hz to 60 Hz

Dimensions

16.9 in x 14.2 in x 14.2 / 43 cm x 36 cm x 36 cm (HxLxP)

Printer

Built-in thermal printer

Throughput

50 tests/hour (Closed Tube) – 60 tests/hour (Open Tube)

Data Management

Complete QC menu, logs (reagents, calibration, maintenance, errors, blank cycle), easy operation

Transmission of patient & QC results to LAN (local area network)

Mono-directional connection

Memory

1000 patients results with histograms

Screen

LCD colour touch screen

Samples

50 tests/hour (Closed Tube) – 60 tests/hour (Open Tube)

Parameters (18)

WBC, RBC, RDW, HGB,  MPV, HCT, MCV, PLT, MCH, MCHC, LYM# & LYM%, MON# & MON%, GRA# & GRA%, PDW*, PCT*

Graphics

PLT, RBC, WBC curves

Blood volume

CBC + 3DIFF mode: 10 μL

Stability

48 hours post-draw stability

Operating sample mode

Open tube or Closed Tube

Identification

Alphanumeric, Integrated barcode reader

 * RUO (Research Use Only)

Results

Parameters

Presicion (cv %)

Linearity

WBC

< 2.5 (at 4 – 10.0 x 103/μL)

0 – 100 x 10³/µL

RBC

< 2.0 (at 4-7 x 106/μL)

0 -8 x 106/µL

HGB

< 1.5 (at 12-18 g/dL)

0 – 26 x g/dL

HCT

< 2.0 (at 36-54 %)

0 – 80%

PLT (whole blood)

< 5.0 (at 200-500 x103/μL)

0 – 2200 x 10³/µL (whole blood)

0 – 4000 x 10³/µL (PLT concentrate)