DiaSys Innova Star

Precision Made Easy

InnovaStar® is an automatic compact POCT analyzer for the daily practice that delivers result quality comparable to measurements on laboratory analyzers.

Easy to use

No special knowledge and extended training is required to operate the system. Testing is made easy with unit-dose reagents and precalibrated tests. The system is calibrated automatically with parameter cards which come with every test package.

Individual hematocrit correction

InnovaStar® uses the individual hematocrit correction to avoid inaccurate calculations of whole blood glucose into plasma glucose. The instrument measures the individual hemoglobin concentration (hematocrit value and Hb concentration are correlated) of a whole blood sample. The measured hemoglobin value is the basis to generate the individual compensation factor for the calculation of whole blood glucose into plasma glucose for each sample – thus providing true glucose values.

Precise

High-quality reagents from DiaSys, freedom from interference, and a photometer detection unit with outstanding linearity are the guarantee for highest possible precision.

Future-Proof

Currently available are tests for HbA1c, glucose and hemoglobin. The future-oriented multiple-testing principle allows for easy expansion of the test portfolio, e.g. for the early detection of kidney disease.

 

Performance data

ParameterHbA1c ISCRP IS
Measuring Range
(units selectable)
3 – 14 %DCCT
9 – 130 mmol/mol
2 – 160 mg/L (Plasma)
5 – 400 mg/L (Whole Blood)
Precision (CV%)<2% at 6.24% DCCT3.1% at 4.11mg/L (Whole Blood)
Sample specimenWhole BloodWhole Blood / Plasma
Sample VolumeDetermination of all analytes from 10 µl whole blood
Measuring Time6.5 minutes to result
Shelf Life18 months18 months

Performance data (continued)

ParameterGlucose/Hemoglobin
GlucoseHemoglobin
Measuring Range
(units selectable)
15 – 800 mg/dL
0.83 – 44.4 mmol/L
5 – 25 g/dL
Precision (CV%)1.3% at 125 mg/L0.79% at 14.5 g/dL
Sample specimenWhole Blood
Sample VolumeDetermination of all analytes from 10 µl whole blood
Measuring Time6.5 minutes to result
Shelf Life18 months
Barcode IdentificationAutomatic barcode scan for reagents; optional barcode scanner for samples
CalibrationPre-calibrated tests
Data Storage50 results
Measurement PrinciplePhotometric / Turbidimetric
Method StorageUp to 19 different methods or lots
EDP InterfaceRS232 connection; USB connection via adapter
Dimensions20 cm (W) x 15 cm (H) x 17 cm (D)
Weight4 kg
Article NumberDescription
1 3329 99 10 760HbA1c IS, 100 tests
2 7069 99 10 760CRP IS, 100 tests
2 7069 99 10 761CRP IS, 50 tests
1 2491 99 10 760Glucose/Hemoglobin IS, 100 tests
97011310 x 100 Sample Cups InnovaStar 10/500
92070910 x 100 Open End capillaries 10 µl
970115300 mL System Solution InnovaStar